Specifications - Sulfur Dioxide Sensors

Sulfur Dioxide

SO2-AE

SO2-AF

SO2-BE

SO2-BF

SO2-D4

Operating range (ppm)

0 to 2,000

0 to 50

0 to 2,000

0 to 100

0 to 20
Sensitivity (nA/ppm) 55 to 80 400 to 650 70 to 85 300 to 480 180 to 360
Resolution (ppm) < 1.5 < 0.1 < 2 < 0.1 < 0.2
Response time (t90) (s) < 30 < 35 < 30 < 40 < 15
Zero current (ppm equivalent) < ± 5 < ± 0.6 < ± 2 < ± 0.5 < ± 0.7
Overgas limit (ppm) 10,000 75 10,000 1,000 50
Operating life (months) > 24 > 24 > 24 > 24 >24*
Temperature range (°C) -30 to 50 -30 to 50 -30 to 50 -30 to 50 -20 to 50
Pressure range (kPa) 80 to 120 80 to 120 80 to 120 80 to 120 80 to 120
Humidity range (%rh) 15 to 90 15 to 90 15 to 90 15 to 90 15 to 90
Load resistor (ohm) recommended 10 to 47 10 to 47 10 to 47 10 to 47 22

Note: all sensors tested and stored at ambient environments unless otherwise stated. As applications of use are outside our control, the information provided is given without legal responsibility. Customers should test under their own conditions, to ensure that the sensors are suitable for their requirements.

*Alphasense will only warrant a maximum continuous exposure period of 10 days at 85% rh and 40°C. Where such exposure occurs the sensor will recover normal electrolyte volumes when allowed to rest at lower %rh and temperature levels for several days.

Please use PCB sockets when connecting Alphasense sensors. Soldering, gluing or other methods of connection will void your warranty.