Specifications - D Miniature Sensors

Miniature Sensors

H2S-D4

CO-D4

NO-D4

NO2-D4

CL2-D4

SO2-D4

HCN-D4

CH-D3
Pellistor

D2

CO
H2S

Operating range (ppm / % LEL)

0 to 100

0 to 1,000

0 to 100

0 to 20

0 to 20

0 to 20

0 to 50

0 to 100

0 to 1,000

0 to 100

Sensitivity (nA/ppm)

110 to 160

30 to 48

500 TO 750

-200 to -400

-200 to -375

180 to 300

25 to 40

11 to 17 mV/%CH4

34 to 55

90 to 150

Resolution (ppm)

< 0.3

< 1.5

< 0.2

< 0.1

< 0.1

< 0.2

< 0.3 -

< 1

< 0.25

Response time (t90) (s)

< 25

< 25

< 20

< 35

< 35

< 12

< 35

< 12

< 25

< 30

Zero current (ppm equiv. @ 20°C)

< ±0.4

< ±3

< ±2

< ±0.35

< ±0.35

< ±0.3

< ±1.5

± 25mV

< ± 6

< ± 0.6

Overgas Limit (ppm)

400

5,000

200

60

60

50

250

 

5,000

400

Operating life (months)*

>24

>24

>24

>24

>24

>24

>12

>24

>24

>24

Temperature range (°C)

-30 to 50

-20 to 50 -20 to 50 -20 to 50 -20 to 50 -20 to 50 -20 to 50 -40 to 50 -30 to 50

Pressure range (kPa)

80 to 120

80 to 120

80 to 120 80 to 120 80 to 120

80 to 120

80 to 120   80 to 120

Humidity range (%rh)

15 to 90

15 to 90

15 to 90 15 to 90 15 to 90

15 to 90

15 to 90   15 to 90

Load resistor (ohm) recommended

10 to 47

10 to 47

33 33 33

22

10 to 47 na 10 to 47

Note: all sensors tested and stored at ambient environments unless otherwise stated. As applications of use are outside our control, the information provided is given without legal responsibility. Customers should test under their own conditions, to ensure that the sensors are suitable for their requirements.

*Alphasense will only warrant a maximum continuous exposure period of 10 days at 85% rh and 40°C. Where such exposure occurs the sensor will recover normal electrolyte volumes when allowed to rest at lower %rh and temperature levels for several days.

Please use PCB sockets when connecting Alphasense sensors. Soldering, gluing or other methods of connection will void your warranty.