Specifications - D Miniature Sensors
|
Miniature Sensors |
H2S-D4 |
CO-D4 |
NO-D4 |
NO2-D4 |
CL2-D4 |
SO2-D4 |
HCN-D4 |
CH-D3 |
D2 |
|
|
CO
|
H2S
|
|||||||||
|
Operating range (ppm / % LEL) |
0 to 100 |
0 to 1,000 |
0 to 100 |
0 to 20 |
0 to 20 |
0 to 20 |
0 to 50 |
0 to 100 |
0 to 1,000 |
0 to 100 |
|
Sensitivity (nA/ppm) |
110 to 160 |
30 to 48 |
500 TO 750 |
-200 to -400 |
-200 to -375 |
180 to 300 |
25 to 40 |
11 to 17 mV/%CH4 |
34 to 55 |
90 to 150 |
|
Resolution (ppm) |
< 0.3 |
< 1.5 |
< 0.2 |
< 0.1 |
< 0.1 |
< 0.2 |
< 0.3 | - |
< 1 |
< 0.25 |
|
Response time (t90) (s) |
< 25 |
< 25 |
< 20 |
< 35 |
< 35 |
< 12 |
< 35 |
< 12 |
< 25 |
< 30 |
|
Zero current (ppm equiv. @ 20°C) |
< ±0.4 |
< ±3 |
< ±2 |
< ±0.35 |
< ±0.35 |
< ±0.3 |
< ±1.5 |
± 25mV |
< ± 6 |
< ± 0.6 |
|
Overgas Limit (ppm) |
400 |
5,000 |
200 |
60 |
60 |
50 |
250 |
|
5,000 |
400 |
|
Operating life (months)* |
>24 |
>24 |
>24 |
>24 |
>24 |
>24 |
>12 |
>24 |
>24 |
>24 |
|
Temperature range (°C) |
-30 to 50 |
-20 to 50 | -20 to 50 | -20 to 50 | -20 to 50 | -20 to 50 | -20 to 50 | -40 to 50 | -30 to 50 | |
|
Pressure range (kPa) |
80 to 120 |
80 to 120 |
80 to 120 | 80 to 120 | 80 to 120 |
80 to 120 |
80 to 120 | 80 to 120 | ||
|
Humidity range (%rh) |
15 to 90 |
15 to 90 |
15 to 90 | 15 to 90 | 15 to 90 |
15 to 90 |
15 to 90 | 15 to 90 | ||
|
Load resistor (ohm) recommended |
10 to 47 |
10 to 47 |
33 | 33 | 33 |
22 |
10 to 47 | na | 10 to 47 | |
|
Note: all sensors tested and stored at ambient environments unless otherwise stated. As applications of use are outside our control, the information provided is given without legal responsibility. Customers should test under their own conditions, to ensure that the sensors are suitable for their requirements. *Alphasense will only warrant a maximum continuous exposure period of 10 days at 85% rh and 40°C. Where such exposure occurs the sensor will recover normal electrolyte volumes when allowed to rest at lower %rh and temperature levels for several days. Please use PCB sockets when connecting Alphasense sensors. Soldering, gluing or other methods of connection will void your warranty. |
||||||||||